The CyberScan TB1000 series turbidity meters (employing
the Nephelometric non-ratio or NTU principle) are designed to meet
stringent requirements of today's scientists. TB1000 series meet the
performance criteria specified by USEPA method 180.1 for NTU measurement
and ISO 7027 (IR version). Ideal for beverage production, drinking
water, waste water treatment, petrochemical and electroplating applications.
Specifications
PRODUCT
SPECIFICATION |
DESCRIPTION |
Turbidity Range |
0 to 1000 NTU |
Resolution
& Accuracy |
0.01 NTU (0 to 9.99
NTU); 0.1 NTU (10 to 99.9 NTU); 1 NTU (100
to 1000 NTU) & ± 2% of Measured Range
+ 0.01 NTU |
Repeatability |
± 1% of Measured
Range + 0.01 NTU |
No. of Calibration
Points |
Up to 3 points |
Response Time |
< 6 seconds for
full step change |
Calibration
Options |
4 points (0.02, 10,
100, 1000) |
Light Source |
White Light (Tungsten)
or Infrared |
Response Time |
< 6 seconds |
Real-time
Clock |
Date and Time |
Output |
RS232 |
Power |
12VDC, using AC adapters
110VAC or 220VAC, 50/60 Hz |
Order Information
ORDER NO.
|
ITEM |
EC-TBDIR1000/10 |
CyberScan TB 1000
Infra-Red Turbidity Bench Meter (NTU) with calibration
kit (ECTBDIRCALKT) & 120 VAC power adapter (EC- 120ADA800) |
EC-TBDIR1000/20 |
CyberScan TB 1000
Infra-Red Turbidity Bench Meter (NTU) with calibration
kit (ECTBDIRCALKT) & 220 VAC power adapter (EC- 220ADA800) |
EC-TBDW1000/10 |
CyberScan TB 1000
White Light Turbidity Bench Meter (NTU) with calibration
kit (ECTBDWCALKT) & 120 VAC power adapter (EC- 120ADA800) |
EC-TBDW1000/20 |
CyberScan TB 1000
White Light Turbidity Bench Meter (NTU) with calibration
kit (ECTBDWCALKT) & 220 VAC power adapter (EC- 220ADA800) |
EC-TBDWCALKT |
Calibration Kit Set
for EC- TBDW1000 (0.02, 10.0, 1000 NTU) |
EC-TBDIRCALKT |
Calibration Kit Set
for EC- TBDIR1000 (0.02, 10.0, 1000 NTU) |
|