ATEX Compliant Intrinsically Safe Pressure Meter, 0 to ±
6900mbar
|
C9507/IS -
> Certified to II 1 G EEx ia IIC T4 Baseefa 03
ATEX 0079, (Certification applicable at atmospheric
pressure)
> High accuracy, electronic pressure meter
> Dust and waterproof to IP67 (BS EN 60529, IEC529)
standards
> Selectable pressure units, PSI, mbar, inH2O,
inHg, mmHg, torr, KPa, cmH2O, Kgcm-2
> Twin inputs for gauge or differential pressure
measurement
> Positive or negative (vacuum) pressure measurement
> Adjustable zero value
> Maximum, minimum pressure and hold function
> Averaged reading function
> Over-range indication - maximum over/under pressure
200PSI/13700mbar
> Selectable auto switch off
> Can function as a calibrator when used with
TK2 pressure test kit
> Supplied with tubing and connectors
> Silicone protection option available for liquid
pressure measurement
ATEX compliant models available from late May 2003 |
Scale
|
Standard
Intrinsically Safe
For Liduids |
C9551
C9501/IS
C9551/SIL |
C9553
C9503/IS
C9553/SIL |
C9555
C9505/IS
C9555/SIL |
C9557
C9507/IS
C9557/SIL |
PSI |
Range
Resolution |
0 to ±2
0.001 |
0 to ±5
0.001 |
0 to ±30
0.01 |
0 to ±100
0.1 |
mbar |
Range
Resolution |
0 to ±140
0.1 |
0 to ±350
0.1 |
0 to ±2000
1 |
0 to ±6900
1 |
inH20 |
Range
Resolution |
0 to ±55
0.01 |
0 to ±140
0.1 |
0 to ±830
0.1 |
0 to ±2750
1 |
inHg |
Range
Resolution |
0 to ±4
0.001 |
0 to ±9.999
0.001 |
0 to ±61
0.01 |
0 to ±200
0.1 |
mmHg |
Range
Resolution |
0 to ±99.99
0.01 |
0 to ±260
0.1 |
0 to ±1500
1 |
0 to ±5200
1 |
torr |
Range
Resolution |
0 to ±99.99
0.01 |
0 to ±260
0.1 |
0 to ±1500
1 |
0 to ±5200
1 |
Pa |
Range
Resolution |
0 to ±9999
1 |
-
- |
-
- |
-
- |
Kpa |
Range
Resolution |
-
- |
0 to ±35
0.01 |
0 to ±200
0.1 |
0 to ±690
0.1 |
mmH20 |
Range
Resolution |
0 to ±1400
1 |
0 to ±3500
1 |
-
- |
-
- |
cmH20 |
Range
Resolution |
-
- |
-
- |
0 to ±1200
1 |
0 to ±7000
1 |
Kgcm2 |
Range
Resolution |
-
- |
-
- |
0 to ±2.1
0.001 |
0 to ±7.0
0.001 |
Maximum
Over Pressure |
6 PSI/400mbar |
15 PSI/1000mbar |
90PSI/620mbar |
200 PSI/13700mbar |
Accuracy
|
0.2%
of full scale at +23°C/+73°C |
Repeatability
|
0.1%
of full scale at +23° C/+73°C |
Mean
Temperature Coefficient
Of Reading |
Better
than 0.1% per °C |
Operation
Range
Temperature
Humidity |
0°C/+32°F
to +50°C/+122°F
10 to 90% RH non condensing |
Environmental
Rating |
IP67,
BS EN 60529, IEC 529 |
Protection
Levels (IS Models only) |
II
1 G EEx ia IIC T4 Baseefa 03 ATEX 0079
(Certification applicable at atmospheric pressure) |
Cennections
|
1/8
BSP female |
Battery
|
9v
PP3 6F22 - MN1604 UCAR9V |
Battery
Life |
90
hours (MN1604) |
|