1 |
Name and chemical formula |
|
The name of the substance to be measured with this
detector tube and
its chemical formula |
|
|
2 |
Detector tube number |
|
The identification number of the detector tube |
|
|
3 |
External appearance |
|
The external appearance of the detector tube(s)
is briefly illustrated. Both
ends of the tube(s) should be broken and prepared before measurements
as shown below. |
|
|
|
|
4 |
Measuring range |
|
The range of the substance concentrations that can
be measured with the detector tube. The standard measuring range,
or range of the printed calibration scale is given in the shaded box.
Some lower limit values of the standard measuring ranges are parenthesized
when they are not printed on the tubes but only their scale lines
are printed. |
|
|
5 |
Number of pump strokes |
|
The number of pump strokes required for measuring
a given range of concentrations. The volume to be sampled with those
strokes is given in the parentheses. The standard pump strokes and
the standard sampling volume are given in the shaded box. With most
Gastec detector tubes, the standard pump strokes are prescribed as
1 (the standard sampling
volume is 100 ml). |
|
|
6 |
Correction factor |
|
When measuring an extended range of concentration,
that is, when the measurement is performed with other than the standard
number of pump strokes, the tube reading should be corrected by multiplying
the reading by the prescribed correction factor. |
|
|
7 |
Sampling time |
|
A waiting time for 100 ml (or 50 ml for half strokes)
to be sampled in the detector tube after the handle of the Model GV-100
Gas Sampling Pump is fully (or half way for 50 ml) pulled out. When
you pump more than once (n times), the waiting time for you to read
the tube, that is, the measuring time is:
Measuring time
= sampling time x n. |
|
|
8 |
Detecting limit |
|
The lower limit of the substance concentration that
the detector tube can detect. This minimal concentration can be recognized
by a slight color change at the entrance of the detecting layer. This
value is followed by the parenthesized number of pump strokes required
for measuring this concentration. |
|
|
9 |
Color change |
|
Indicates how the detecting layer of the detector
tube will change its color by the
reaction with the target substance. This is shown in the form of:
Original color ---> Changed color |
|
|
10 |
Reaction principle |
|
The chemical reaction of the target substance with
the reagent(s) in the
detector tube is briefly stated. |
|
|
11 |
Coefficient of Variation |
|
An indicator of the accuracy of the detector tube.
It represents how the
tube indications may deviate from their mean value in percentage:
Standard
deviation
Coefficient
of Variation = ------------------------- * 100
Mean
value
|
|
|
12 |
Shelf life |
|
The period that Gastec will warrant the quality
of the detector tubes, provided that they are stored under the prescribed
conditions |
|
|
13 |
Corrections for temperature & humidity
|
|
For detector tubes whose indications are affected
+/- 10 % or more by the fluctuation of temperature or humidity, the
tube readings should be corrected. Whether or not such corrections
are necessary are indicated here. The correction data, if necessary,
is provided in the instruction sheets provided with the detector tubes.
|
|
|
14 |
Storage condition |
|
To keep the high quality of Gastec detector tubes,
it is necessary to store them in a cool and dark place. It is recommended
to store the tubes in the refrigerator, i.e, 0 to 10oC(32 to 50oF)
even if the tube labels describes cool and dark place. |
|
|
15 |
Possible coexisting substances and their
interferences |
|
Substances that are liable to coexist with the target
substance and their influences on the tube indications are listed
here. The column "Concentration" Iists the lowest concentration
that may affect the indication +/-10 % or more, or the highest concentration
that is assured of no influence to such extent. These concentration
levels are expressed either as quantitative ratios of interferents
to the target substance (e.g., 1/5, 2 times)or concentrations of interferents
themselves (e.g., 3%, 500ppm).Where, reads"equal to
or higher than" and reads "equal to or lower than"
.The presumed influence by coexistence is given in the column "Interference"
where the " + " sign suggests 10% or higher indication,
the " - " sign expresses - 10% or lower indication, or "No"means
noinfluence. "Bleaching" means that this interferent will
cause the reaction colorto be pale. For your reference, the column
"Changes color by itself to" is givento show how the interferents
will change the color of the detecting layer if they exist without
the presence of the target gas. "No" represents that no
color change will be observed. The table of this interference gases
primarily expresses the interference of each coexisting gas in the
gas concentration range, equivalent to the gas concentration by some
of the other coexisting gases or vapors mentioned from the main purposes
of using this detector tube. Therefore, the test result may be given
positive result by the other substance not listed in the table. For
more precise
information is needed, please contact us or our distributors in your
territory. |
|
|
16 |
Other substance (s) measurable with this
detector tube |
|
Some detector tubes have high sensitivities to several
other substances and can be used to measure those concentrations.
If this is the case, the names of such substances, the correction
methods of the tube readings (whether by using a correction scale
or a correction factor), the numbers of pump strokes and the measuring
ranges are listed. For detector tubes that perform qualitative analysis,
the title of this item is read as "Substances
measurable with this detector tube." |
|
|
17 |
Calibration gas generation |
|
The method of generating the calibration gas that
is used for calibrating
the detector tube or testing its accuracy. |
|
|
18 |
Threshold limit values and explosive range |
|
Some substances may be hazardous to health or dangerous
to life depending on their concentrations or the duration of exposure,
or may be combustible or explosive. For detector tubes for measuring
those substances, the threshold limit values based upon the recommendations
of the American Conference of Governmental Industrial Hygienists
(ACGIH), as well as the explosive ranges are presented for your reference.
TLV-TWA :Threshold Limit Value-Time-Weighted
Average The time-weighted average concentration for a normal 8-hour
workday or 40-hour workweek, to which nearly all workers may be repeatedly
exposed, day after day, without adverse effect.
TLV-STEL:
Threshold Limit Value-Short-Term Exposure Limit The maximum concentration,
to which workers can be exposed for a period of not longer than 15
minutes continuously without suffering from:
( I ) irritation,
( 2 ) chronic or irreversible tissue damage, or
( 3 ) narcosis of sufficient degree to increase the likelihood of
accidental injury, impair self-rescue, or materially reduce work efficiency,
and provided that the daily TLV-TWA is not exceeded.
TLV-C:Threshold
Limit Value-Ceiling
The concentration that should not be exceeded during any part of the
working exposure. In conventional industrial hygiene practice if instantaneous
monitoring is not feasible, the TLV-C can be assessed by sampling
over a 15 minute period except for those substances that may cause
immediate irritation when exposures are short.
Explosive
range :
The range of concentrations of the substance in which its gas
or vapor mixed with air will be explosive. Concentrations are
expressed by the percentage ( % ) of the substance in the
mixed air. |