1,3-Butadiene CH2:CHCH:CH2 NO.174D
 Performance
Measuring Range
1.3 to 200 ppm
Sampling Hours
1 to 8 hours
Detecting Limit
1.3 ppm (8 hours)
Color Change
Reddish purple ---> Pale brown
Reaction Principle Butadiene reacts with permanganate to produce light brown reaction product.
Coefficient of Variation 10% (for 10 to 200 ppm¥hr)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.

Possible coexisting substances and their interferences
Substance
 Interference Change color by itself
 Unsaturated hydrocarbons
Plus error
Produce pale brown discoloration

Other substance measurable with this detector tube
Substance
 Correction Factor
Measuring Range
 Ethylene
1.2
1.5 to 240 ppm
 Vinyl chloride
1.2
1.5 to 240 ppm
 Isoprene
2.0 2.5 to 400 ppm
 Trans- 1,2- Dichloroethylene
3.0 3.8 to 600 ppm

Calibration gas generation Permeation tube method

TLV-TWA
TLV-STEL
Explosive range
2ppm - 2.0 to 12%