Hydorogen chloride HCl NO.14D
 Performance
Measuring Range
1 to 100 ppm
Sampling Hours
1 to 10 hours
Detecting Limit
0.05 ppm (10 hours)
Color Change
Yellow ---> Purple
Reaction Principle Hydrogen chloride react with base to discolor the indicator to purple.
Coefficient of Variation 10% (for 10 to 100 ppm-hr)
Shelf Life 3 Years
Corrections for temperature & humidity Temperature correction is necessary
Humidity correction is necessary
Store the tubes at cool and dark place.

 Possible coexisting substances and their interferences
Substance
 Concentration
 Interference Change color by itself
 Chlorine
1/5 times
Plus error
(Bleaches zero zone)
Bleaches zero zone
 Nitric acid
1/5 times
Plus error
Produce purple stain
 Hydrogen fluoride
1 time
Plus error Produce purple stain

Other substance measurable with this detector tube
Substance
 Correction Factor
Measuring Range
 Nitric acid
0.8
0.8 to 80 ppm
 Hydrogen fluoride
2.5
2.5 to 250 ppm

Calibration gas generation Diffusion tube method

TLV-TWA
TLV-STEL
Explosive range
- C 5ppm -