The detector uses a silicon photodiode sensor to reduce size and prolong working life.
All the necessary features are built right into a single rack-sized unit, including a reference-gas generator, an ozone-source drier unit, an ozone decomposer, and a sampling pump. No supplemental gas is required.
Principles
Cross flow modulation type, reduced pressure chemiluminescence (CLD)
The chemiluminescence method uses the reaction of NO with O3
NO+O3 → NO2*+O2
NO2+NO2+h ν
A portion of the NO2 generated as the result of this reaction becomes NO2*. As these excited molecules return to the ground state, chemiluminescence is generated in the range of 600 nm to 3,000 nm. The light intensity is in proportion to the concentration of NO molecules and by measuring it we obtain the NO concentration of the sample. A deoxidation conver-ter changes the NO2 to NO, which is measured. In other words, the NO2 concentration can be obtained by the difference between (1) the NOx concentration measured when the sample gas is directed through a converter and (2) the NO concentration measured when the gas is not run through the converter.
Specifications
PRODUCT SPECIFICATION |
DESCRIPTION |
Principle: |
Cross flow modulation type, reduced pressure chemiluminescence (CLD) |
Application: |
NO2, NO and NOx in ambient air |
Range:
|
Standard ranges: 0-0.1/0.2/0.5/1.0 ppm; auto range ~ manual range selectable;
can be operated by remote switching.
Optional (measurable) ranges: 4 ranges selectable from 0-10 ppm, within 10 times range ratio;
auto range ~ manual range selectable; can be operated by remote switching. |
Repeatability:
|
±1.0% F.S. |
Linearity: |
±1.0% F.S |
Sample gas flow rate: |
Approx. 0.8L/min |
Indication: |
Measured value, range, alarm, maintenance screen |
Alarms: |
During AIC, zero calibration error, span calibration error, temperature error in converter, etc. |
On-screen messages are available in four languages: |
English, German, French, and Japanese. |
Input/output: |
• 0-1 V/0-10 V/4-20 mA, to be specified (2 systems: either (1) momentary value and integrated or (2) moving average value)
• Contact input/output
• RS-232C |
Power: |
100/110/115/120/220/230/240 VAC, 50/60 Hz (to be specified) |
Dimensions: |
430(W) x 550(D) x 221(H) mm |
Mass: |
Approx. 21 kg |
|