  Products > Ambient Air Quality Monitors Stations (AAQMS) > Sub Products > Ambient CO Monitor

Features
 
The cross flow modulation type, infrared-absorption technology eliminates the need for adjusting optical alignment.
For the user, this means very stable and sensitive (5 ppm F.S.) measurements.

The APMA-370 uses an AS-type interference-compensating detector, and a purified reference gas. The reference gas is generated by purging the sample through an oxidation process, where an oxidizing catalyst burns the CO to CO2. These features eliminate interference from other elements, resulting in highly accurate measurements.
The APMA-370 does not use such components as reflecting mirrors, that attract foreign matter. This means the optical bench stays clean assuring you of stable results over long periods of time.

Principle

Cross flow modulation, inflared (NDIR) absorption technology

Conventional technology uses an optical chopper to obtain modulation signals. Instead, the APMA-370 uses solenoid valve cross flow modulation. Fixed amounts of the sample gas and the reference gas are injected alternately into the measurement cell. With the cross flow-modulation method, if the same gas is used for both the sample gas and the refe-rence gas (e.g., zero gas could be used for both), no modulation signal will be generated. This has the great advantage that, in principle, when analyzing minute amounts of gas there is no generation of zero-drift. An additional advantage is that the elimi-nation of rotary sectors precludes the need for optical adjustment. These features assure greatly improved stability over long periods of measurement. A further improvement is that in the front chamber of the detector, the measurable components, including interference components, are detected; in the rear chamber, only interference components are detec-ted. By means of subtraction processing, the actual signal obtained is one that has very little interference.


Specifications

 PRODUCT SPECIFICATION  DESCRIPTION
 Principle:  Cross flow modulation, non-dispersive  infrared (NDIR) absorption technology
 Application:  CO in ambient air
 Range:
 Standard ranges: 0-10/20/50/100 ppm;  0- 5/10/20/50 ppm; auto range ~ manual  range selectable; can be operated by  remote switching.
 Optional (measurable) ranges: 4 ranges  selectable from 0-100 ppm, within 10 times  range ratio; auto range ~ manual range  selectable; can be operated by remote  switching.
 Repeatability:
 ±1.0% F.S.
 Linearity:  ±1.0% F.S
 Sample gas flow rate:  Approx. 1.5L/min
 Indication:  Measured value, range, alarm,  maintenance screen
 Alarms:  During AIC, zero calibration error, span  calibration error, temperature error in  catalyzer, etc.
 On-screen messages are  available in four  languages:  English, German, French, and Japanese.
 Input/output:  • 0-1 V/0-10 V/4-20 mA, to be specified (2  systems: either (1) momentary value and  integrated or (2) moving average value)
 • Contact input/output
 • RS-232C
 Power:  100/110/115/120/220/230/240 VAC, 50/60  Hz (to be specified)
 Dimensions:  430(W) x 550(D) x 221(H) mm
 Mass:  Approx. 16 kg

Dimension Outline